Test Code STFR Soluble Transferrin Receptor (sTfR), Serum
Reporting Name
Soluble Transferrin Receptor (sTfR)Useful For
Evaluation of suspected iron deficiency in patients who may have inflammation, infection, or chronic disease and other conditions in which ferritin concentration does not correlate with iron status, including:
-Cystic fibrosis patients who frequently have inflammation or infections(2-3)
-Evaluating insulin-dependent diabetics who may have iron-deficiency resulting from gastric autoimmunity and atrophic gastritis(4)
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SerumSpecimen Required
Supplies: Sarstedt Aliquot Tube 5 mL (T914)
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Serum gel tubes should be centrifuged within 2 hours of collection.
2. Red-top tubes should be centrifuged, and the serum aliquoted into a plastic vial within 2 hours of collection.
Specimen Minimum Volume
0.25 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Frozen (preferred) | 90 days | |
Refrigerated | 7 days | ||
Ambient | 72 hours |
Reference Values
1.8-4.6 mg/L
It is reported that African Americans may have slightly higher values.
Day(s) Performed
Monday through Sunday
Test Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
84238
Clinical Information
Iron uptake into cells is mediated through internalizing iron-transferrin complexes. The iron-transferrin complex binds to transferrin receptors present on the external face of the plasma membrane and is internalized through endosomes with ultimate release of iron into the cytoplasm. Plasma membrane-bound transferrin receptor is released by proteolytic cleavage of the extracellular domain, resulting in the formation of a truncated soluble transferrin receptor (sTfR) that circulates freely in the blood.
The concentration of sTfR is an indicator of iron status. Iron deficiency causes overexpression of transferrin receptor and sTfR levels, while iron repletion results in decreased sTfR levels. While ferritin measurement is the accepted method for assessment of iron deficiency, ferritin is an acute-phase reactant and elevates in response to processes that do not correlate with iron status, including inflammation, chronic disease, malignancy, and infection. sTfR is not an acute-phase reactant and the interpretation of iron status using sTfR measurement is not affected by these confounding pathologies.(1)
Report Available
Same day/1 to 2 daysReject Due To
Gross hemolysis | Reject |
Method Name
Immunoturbidimetric Assay