Sign in →

Test Code PNRCH Drug Immunoassay Panel, Urine

Reflex Tests

Test ID Reporting Name Available Separately Always Performed
BARBU Barbiturates Confirmation, U Yes No
COKEU Cocaine and metabolite Conf, U Yes No
THCU Carboxy-THC Confirmation, U Yes No

Method Name

Only orderable as part of profile. For more information see:

-CSMPU / Controlled Substance Monitoring Panel, Random, Urine.

-ADMPU / Addiction Medicine Profile with Reflex, 22 Drug Classes, High Resolution Mass Spectrometry and Immunoassay Screen, Random, Urine

-CSMEU / Controlled Substance Monitoring Enhanced Profile with Reflex, 21 Drug Classes, High Resolution Mass Spectrometry and Immunoassay Screen, Random, Urine

 

Immunoassay

Reporting Name

Drug Immunoassay Panel, U

Specimen Type

Urine


Ordering Guidance


The test does not screen for drug classes other than those listed in Testing Algorithm.



Specimen Required


Only orderable as part of profile. For more information see:

-CSMPU / Controlled Substance Monitoring Panel, Random, Urine

-ADMPU / Addiction Medicine Profile with Reflex, 22 Drug Classes, High Resolution Mass Spectrometry and Immunoassay Screen, Random, Urine

-CSMEU / Controlled Substance Monitoring Enhanced Profile with Reflex, 21 Drug Classes, High Resolution Mass Spectrometry and Immunoassay Screen, Random, Urine


Specimen Stability Information

Specimen Type Temperature Time Special Container
Urine Refrigerated (preferred) 14 days
  Frozen  14 days
  Ambient  72 hours

Reference Values

Only orderable as part of profile. For more information see:

-CSMPU / Controlled Substance Monitoring Panel, Random, Urine

-ADMPU / Addiction Medicine Profile with Reflex, 22 Drug Classes, High Resolution Mass Spectrometry and Immunoassay Screen, Random, Urine

-CSMEU / Controlled Substance Monitoring Enhanced Profile with Reflex, 21 Drug Classes, High Resolution Mass Spectrometry and Immunoassay Screen, Random, Urine

 

Negative

Screening cutoff concentrations:

Barbiturates: 200 ng/mL

Cocaine (benzoylecgonine-cocaine metabolite): 150 ng/mL

Tetrahydrocannabinol carboxylic acid: 50 ng/mL

 

This report is intended for use in clinical monitoring or management of patients. It is not intended for use in employment-related testing.

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information

80307

Useful For

Detecting drug use involving barbiturates, cocaine, and carboxy-tetrahydrocannabinol

 

This test is not intended for use in employment-related testing.

Specimen Minimum Volume

20 mL

Reject Due To

  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability

Clinical Information

This test uses the simple screening technique that involves immunoassay testing for drugs by class. All positive immunoassay screening results are confirmed by either gas chromatography mass spectrometry (GC-MS) or liquid chromatography tandem mass spectrometry (LC-MS/MS) and quantitated before a positive result is reported.

 

This assay was designed to test for and confirm by GC-MS the following:

-Barbiturates

-Cocaine

 

This assay was designed to test for and confirm by LC-MS/MS the following:

-Carboxy-tetrahydrocannabinol

 

This test is intended to be used in a setting where the test results can be used to make a definitive diagnosis.

Day(s) Performed

Monday through Saturday

Report Available

Same day/1 to 2 days