Test Code CRCOF Chromium and Cobalt, Synovial Fluid
Ordering Guidance
This test should only be used in individuals with chromium or cobalt implants.
Specimen Required
Patient Preparation: High concentrations of gadolinium and iodine are known to interfere with most inductively coupled plasma mass spectrometry-based metal tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for at least 96 hours.
Supplies: Metal Free B-D Tube (EDTA), 6 mL (T183)
Container/Tube: Royal blue top (metal-free EDTA)
Specimen Volume: 1 mL
Collection Instructions: See Metals Analysis Specimen Collection and Transport for complete instructions.
Additional Information: Cobalt and chromium are present in the black rubber plunger seals found in most disposable syringes. As a result, synovial fluid should not be collected in these devices as contamination may occur.
Useful For
Monitoring metallic prosthetic implant wear and local tissue destruction in failed hip arthroplasty constructs
This test is not useful for assessment of vitamin B12 activity.
Profile Information
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
CRSY | Chromium, Synovial Fl | Yes | Yes |
COSY | Cobalt, Synovial Fl | Yes | Yes |
Special Instructions
Method Name
Inductively Coupled Plasma Mass Spectrometry (ICP-MS)
Reporting Name
Chromium and Cobalt, Synovial FlSpecimen Type
Synovial FluidSpecimen Minimum Volume
0.4 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Synovial Fluid | Refrigerated (preferred) | 90 days | |
Ambient | 90 days | ||
Frozen | 90 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Clinical Information
Per US Food and Drug Administration recommendations, orthopedic surgeons should consider measuring and following serial chromium and cobalt concentrations in EDTA anticoagulated whole blood in symptomatic patients with metal-on-metal hip implants as part of their overall clinical evaluation. However, a recent publication(1) has shown synovial fluid measurements were superior to whole blood and serum chromium and cobalt concentrations in predicting local tissue destruction in failed hip arthroplasty constructs.
Prosthetic devices produced by Depuy Company, Dow Corning, Howmedica, LCS, PCA, Osteonics, Richards Company, Tricon, and Whiteside are typically made of chromium, cobalt, and molybdenum. This list of products is incomplete, and the products' compositions change occasionally; see each prostheses' product information for composition details.
Chromium:
Chromium (Cr) is a naturally occurring element widely distributed in the environment. It exists in several valence states with the 3 main forms being Cr(0), Cr(III), and Cr(VI). Cr(III) is an essential trace element that enhances the action of insulin. Deficiency leads to impaired growth, reduced life span, corneal lesions, and alterations in carbohydrates, lipid, and protein metabolism.
Chromium is widely used in manufacturing processes to make various metal alloys, such as stainless steel. It is also used in many consumer products, including wood treated with copper dichromate, leather tanned with chromic sulfate, and metal-on-metal hip replacements.
The general population is most likely to be exposed to trace levels of chromium, as Cr(III) is naturally occurring in foods, such as fruits, vegetables, nuts, beverages, and meats. The highest potential occupational exposure occurs in the metallurgy and tanning industries, where workers may be exposed to high air concentrations.
Cobalt:
Cobalt is a naturally-occurring, hard, gray element widely distributed in the environment. It is used to produce alloys in the manufacturing of aircraft engines, cutting tools, and some artificial hip and knee joint prosthesis devices.
Cobalt is an essential cofactor in vitamin B12, which is necessary for neurological function, brain function, and the formation of blood. For most people, food is the largest source of cobalt intake. However, more than a million workers are potentially exposed to cobalt and its compounds, with the greatest exposure in mining processes, cemented tungsten-carbide industry, cobalt powder industry, and alloy production industry.
Cobalt is not highly toxic, but large doses will produce adverse clinical manifestations. Acute symptoms include pulmonary edema, allergy, nausea, vomiting, hemorrhage, and kidney failure. Chronic exposure to cobalt-containing hard metal (dust or fume) can result in a serious lung disease called hard metal lung disease, which is a type of pneumoconiosis (lung fibrosis). Furthermore, inhalation of cobalt particles can cause respiratory sensitization, asthma, shortness of breath, and decreased pulmonary function. Even though the primary route of occupational exposure to cobalt is the respiratory tract, skin contact is also important because dermal exposures to hard metal and cobalt salts can result in significant systemic uptake. Sustained exposures can cause skin sensitization, which may result in eruptions of contact dermatitis. In cases of suspected toxicity, blood, serum, or urine concentrations of cobalt can be checked. Vitamin B12 should be used to assess nutritional status.
Reference Values
CHROMIUM:
0-17 years: Not established
≥18 years: <16.9 ng/mL
COBALT:
0-17 years: Not established
≥18 years: <19.8 ng/mL
Day(s) Performed
Friday
Report Available
2 to 8 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
83018
82495