Test Code CIFA Cutaneous Immunofluorescence Antibodies, IgA, Serum
Specimen Required
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 2 mL
Collection Instructions: Centrifuge and aliquot serum into a plastic vial.
Useful For
Confirming the presence of IgA antibodies to diagnose pemphigoid, pemphigus, epidermolysis bullosa acquisita, or bullous lupus erythematosus
Method Name
Indirect Immunofluorescence Assay (IFA)
Reporting Name
Cutaneous Immfluor. Ab (IgA), SSpecimen Type
SerumSpecimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 14 days | |
Frozen | 30 days | ||
Ambient | 14 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | Reject |
Gross icterus | OK |
Clinical Information
Immunoglobulin A anti-basement membrane zone (BMZ) antibodies are produced by patients with pemphigoid. In most patients with bullous pemphigoid, serum contains IgA anti-BMZ antibodies, while in cicatricial pemphigoid circulating IgA anti-BMZ antibodies are found in a minority of cases. Sensitivity of detection of anti-BMZ antibodies is increased when serum is tested using sodium chloride-split primate skin as substrate.
Circulating IgA anti-BMZ antibodies are also detected in patients with epidermolysis bullosa acquisita and bullous eruption of lupus erythematosus.
IgA anti-cell surface (CS) antibodies are produced by patients with pemphigus. The titer of anti-CS antibodies generally correlates with disease activity of pemphigus.
Reference Values
Report includes presence and titer of circulating antibodies. If serum contains basement membrane zone antibodies on split-skin substrate, patterns will be reported as:
1) Epidermal pattern, consistent with pemphigoid
2) Dermal pattern, consistent with epidermolysis bullosa acquisita
Negative in normal individuals
Day(s) Performed
Monday through Friday
Report Available
2 to 7 daysPerforming Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
88346
88350